Appendix A: Data Format
Appendix-2 Pulse Oximeter for Sleep Screening Operation Manual
The following table shows an example of data de-compression:
Byte in HEX De-compression
88 Marker byte, expect the next byte to be SpO2;
60 SpO2 value, HEX 60 equals to 96% SpO2 decimal, expect the next byte
to be Pulse Rate;
52 Pulse Rate, HEX 52 equals to 82 bpm decimal, expect the next byte to be
the byte divided into low and high nibbles;
f0 Low nibble is 0, SpO2 = 96% + 0 = 96%;
High nibble is –1, Pulse Rate = 82bpm – 1 = 81 bpm;
ef Low nibble is -1, SpO2 = 96% - 1 = 95%;
High nibble is –2, Pulse Rate = 82bpm – 2 = 80 bpm;
30 Low nibble is 0, SpO2 = 95% + 0 = 95%;
High nibble is 3, Pulse Rate = 80bpm + 3 = 83 bpm;
Note: Your service representative can supply an ANSI C de-compression routine.
Appendix B: Setting the Optional Real Time Clock
1. Clear trend memory by pressing and holding the print key at power up until CLR
stops flashing.
2. Turn the monitor off. Power on the monitor and continue to hold the on key for 6-8
seconds. The version number should be displayed and then the display will go blank
except for one set of digits. This is the minutes setting.
3. Adjust the minutes by pressing the print key until the desired setting is reached.
4. Press the power key again to show the hour setting. Use the print key to set the
desired hour.
5. Press the power key again to show the day setting. Use the print key to set the
desired day.
6. Press the power key again to show the month setting. Use the print key to set the
desired month.
7. Press the power key again to show the year setting. Use the print key to set the
desired year.
8. Press the power key again to exit the time setting menu. After the power key is
pressed a time stamp is printed and the monitor turns itself off.
9. Verify that the correct time stamp was printed. If the correct time and date was not
printed, repeat steps 2 through 9 until the proper time and date is verified.
The time and date can be verified by repeating step 2 and then pressing the print key until
the current time and date is printed and the monitor turns itself off.