
Chapter 1: Introduction
Pulse Oximeter for Sleep Screening Operation Manual 1-1
Chapter 1: Introduction
About the Manual
The Operation Manual provides installation, operation, and maintenance instructions for
health-care professionals and other users, trained in monitoring respiratory and cardiovascular
activity.
Theory of Operation
The oximeter determines SpO
2
and pulse rate by passing two wavelengths of low intensity
light, one red and one infrared, through body tissue to a photodetector. During measurement,
the signal strength resulting from each light source depends on the color and thickness of the
body tissue, the sensor placement, the intensity of the light sources, and the absorption of the
arterial and venous blood (including the time varying effects of the pulse) in the body tissues.
Figure 1.0: Theory of Operation
1. Red and infrared LED light sources
2. Detector
The oximeter processes these signals, separating the time invariant parameters (tissue
thickness, skin color, light intensity, and venous blood) from the time variant parameters
(arterial volume and SpO
2
) to identify the pulse rate and calculate oxygen saturation. Oxygen
saturation calculations can be performed because oxygen saturated blood predictably absorbs
less red light than oxygen depleted blood.
Since measurement of SpO
2
depends on a pulsating vascular bed, any condition that restricts
blood flow, such as use of a blood pressure cuff or extremes in systemic vascular resistance,
may cause an inability to determine accurate pulse and SpO
2
readings.
Commentaires sur ces manuels